Teardown and repair of an Agilent E3632A DC power supply

pE3632A-e1505951806733-600

Teardown and repair of an Agilent E3632A DC power supply from The Signal Path:

In this episode Shahriar & Rosanah investigate an Agilent power supply which does not appear to power on. It can be quickly observed that the fuse has failed on the unit. Using an isolation transformer a small amount of AC voltage is applied to the unit after the fuse replacement. It is clear that a short is present somewhere in the instrument since even at 10V AC the instrument consumes more than 1A.

See the full post at thesignalpath.com.

Check out the video after the break.